| Catalog Number |
SXMP1B408 |
| Medical Device Problem Code |
Failure To Adhere Or Bond (1031)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2018 and barbed suture was used.During the procedure, the suture was not holding.The suture appears to not have any barbs on it.It is unknown how the procedure was completed.There were no adverse consequences for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > pc-000199791 date sent to the fda: 07/12/2018 the device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Investigation summary = > a needle-suture piece and one top foil were returned for evaluation.During the visual inspection of the opened sample, the swage and attachment area were as expected and a small suture piece still was attached to barrel needle.In addition, the end of the suture was cut appears to be by use of the surgical instrument and barbs could not be measured.No conclusion could be reached as to what may have caused the reported incident.Unopened samples were also returned for analysis.In the visual inspection of the unopened samples, no defects were found on the packages.The samples were opened and the swage and attachment area of the needles were as expected.The suture was examined and the barbs were present along of the strand.Also, 10 barbs were randomly measured and all barbs met with the requirements.In addition, the loops were as expected.According to the condition of the samples, the barbs met the specification.
|
| |
|
Search Alerts/Recalls
|