Product complaint #: (b)(4).The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Investigation summary: a dispensed needle-suture combination, an empty labeled winding former, one empty opened foil, two top foils and unopened samples were returned for analysis.During the visual inspection of the opened sample, the swage and attachment area were as expected.The suture was examined and the barbs were present along of the strand; also, 10 barbs were randomly measured and all barbs met with the requirements.In addition, the loops were as expected.In the visual inspection of the unopened samples, no defects were found on the packages.The samples were opened and the swage and attachment area of the needles were as expected.The suture was examined and the barbs were present along of the strand.Also, 10 barbs were randomly measured and all barbs met with the requirements.In addition, the loops were as expected.According to the condition of the samples, the barbs met the specification.
|