| Model Number |
48753212 |
| Medical Device Problem Code |
Labelling, Instructions for Use or Training Problem (1318)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while in the or, a cage was loaded onto the inserter.It was noticed by the surgeon that the anterior / posterior markers were on the wrong/opposite sides of the cage.There are no reports of adverse consequences to the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that while in the or, a cage was loaded onto the inserter.It was noticed by the surgeon that the anterior / posterior markers were on the wrong/opposite sides of the cage.There are no reports of adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: a video clip device history review, complaint history review, risk assessment.Results/conclusion: the reported laser marking error was confirmed via a video clip of the device sent to post market surveillance.The device was not returned.The event did not result in any patient's harm.A non conformance was opened to further investigate this issue.A commercial hold has been placed on the affected items and the product in the warehouse was scraped.A change control has been open to add a laser marking inspection before product release.Capa was open to evaluate how to avoid a reoccurrence of this issue and to continue the investigation regarding the laser marking process.
|
| |
|