WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA HELICAL BLADE 75MM STERILE; ROD, FIXATION, INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
04.038.275S |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient weight is unknown.Date of device migration is unknown.Date of implantation is unknown.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.A device history records review has been requested.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a hardware removal and revision on (b)(6) 2018, due to the helical blade cutting into the femoral head.The patient was initially implanted with one (1) trochanteric fixation nail advanced (tfna) nail, one (1) tfna helical blade, 1.55mm cortex screw and a distal locking screw on an unknown date.Post-operatively, the patient started having hip pain and it was noticed, with the help of x-rays, that the helical blade had migrated into the femoral head.The patient was brought back to the operating room and was revised to a new tfna on (b)(6) 2018.The procedure was completed successfully with no reported delay, and the patient was reported to be doing fine after the surgery.Subsequently, after the first revision procedure, the patient started experiencing hip again and was brought back to the operating room on (b)(6) 2018.This complaint only captures the first revision that was completed on (b)(6) 2018.The patient had the second revision completed on (b)(6) 2018 and is captured in (b)(4).Concomitant devices: 10mm/130 degree titanium cannulated tfna nail 340mm/left ¿ sterile (part: 04.037.055s, lot: 9814022, qty: 1), 1.55mm cortex screw (part: 02.214.122, lot: unknown, qty: 1), distal locking screw (part: unknown lot: unknown, qty: 1).This report is for a tfna helical blade 75mm sterile.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device history record (dhr) review: part number: 04.038.275s; lot number: 7968806; date of manufacture: 13 april 2015; place of manufacture: (b)(4); part expiration date: 31 march 2025; nonconformance noted: n/a.Description of dhr review: a review of the device history record revealed no complaint related anomalies.The device history record shows this lot of tfna helical blade 75mm sterile product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this lot met all specifications with no non-conformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Investigation summary: investigation flow: device interaction/functional.Visual inspection: visual inspection performed at customer quality (cq) showed wear consistant with implantation and explantation.However, no damage that would contribute to the reported complaint condition was observed.The complaint device was received with the concomitant nail which had no allegation and showed no evidence of contributing to the reported event.No x-rays/radiographs showing the reported migration condition were provided, therefore this complaint was not able to be confirmed.Functional test: a relevant functional test to replicate the reported complaint condition was not able to be performed at cq because this was a post operative event.Document / specification review: the complaint device was manufactured in april 2015.A review of the device history record revealed no complaint related anomalies.The device history record shows this lot of tfna helical blade 75mm sterile product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this lot met all specifications with no non-conformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Tabulated product design drawing for the family of titanium trochanteric fixation nail advanced (tfna) screws was reviewed during this investigation.No product design issues or discrepancies were observed.Dimensional inspection: the outside diameter of the distal major threads measured ø10.24mm at cq (ca592) which is within specification of ø10.3mm +0.0/-0.1mm per tabulated product design drawing.Conclusion: a root cause for the reported migration was not able to be determined based on the provided information.This complaint was not able to be confirmed based on the provided information, no product design issues or manufacturing discrepancies that would contribute to the reported complaint condition were identified during this investigation, the calculated occurrence rate does not exceed the estimated rate in the dcrm and no new patient harms were identified as a result of this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|