Model Number TRUEMETRIX
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Information (3190)
Date of Event 05/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4) product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Most likely underlying root cause: mlc-14 - insufficient blood sample applied to strip (user) test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Unable to send product letter.No address on file for the customer.
 
Event or Problem Description
Consumer reported complaint for e-2 accompanied by symptoms.The expected fasting blood glucose test result range is undisclosed.The customer did report symptom of feeling sick.Customer can't describe the symptoms.Medical attention is not reported as a result of the actual blood glucose results and symptoms.During the call on (b)(6) 2018, a blood test was performed by customer and produced test results of 249mg/dl using true metrix meter.The product is not stored by customer according to specification in the bathroom.The test strip lot manufacturer's expiration date is 06/22/2019 and open vial date is unknown.The meter memory was not reviewed for previous test result history.During the call meter read the blood 249mg/dl.Customer does not know if this is a good number for her or not.Customer was advised to speak to her doctor about her glucose readings and advised to call 911 or her doctor if her symptoms persist.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7616022
Report Number1000113657-2018-00663
Device Sequence Number1695166
Product Code NBW
UDI-Device Identifier00010939722447
UDI-Public(01)00010939722447
Combination Product (Y/N)N
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2724
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2018
Initial Report FDA Received Date06/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-