| Model Number |
6248V-130 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during an implant procedure, the catheter broke midway down and completely (360 degrees around.) the catheter was cut with scissors.A second catheter was tried, but the same issue occurred and additionally the deflectable wire came out while the catheter was being slit.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the delivery catheter was returned and analyzed.The mechanical operation of the catheter peeling/slitting/splitting showed a spiral slit.The mechanical operation of the catheter shaft was bent.Visual analysis of the catheter indicated damage during use.If information is provided in the future, a supplemental report will be issued.
|
| |
|