Model Number 6248V-130
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during an implant procedure, the catheter broke midway down and completely (360 degrees around.) the catheter was cut with scissors.A second catheter was tried, but the same issue occurred and additionally the deflectable wire came out while the catheter was being slit.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product event summary: the delivery catheter was returned and analyzed.The mechanical operation of the catheter peeling/slitting/splitting showed a spiral slit.The mechanical operation of the catheter shaft was bent.Visual analysis of the catheter indicated damage during use.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN SELECT II
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7616084
Report Number9612164-2018-01491
Device Sequence Number16693058
Product Code DQY
UDI-Device Identifier00613994450753
UDI-Public00613994450753
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K061480
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/26/2020
Device Model Number6248V-130
Device Catalogue Number6248V-130
Device Lot Number0009050442
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Initial Date Received by Manufacturer 05/30/2018
Supplement Date Received by Manufacturer09/13/2018
11/19/2018
Initial Report FDA Received Date06/19/2018
Supplement Report FDA Received Date09/17/2018
11/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
-
-