DATASCOPE MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Crack (1135); Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge field service engineer (fse) was dispatched to the customer's site.The fse replaced the lcd display and completed the repair with full calibration, functional testing and safety checks to factory specifications.The iabp passed all functional and safety tests per factory specifications, returned to customer and cleared for clinical use.
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) monitor lcd screen was cracked and not working.It is unknown under which circumstances this event occurred.However, no patient involvement or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7616622
MDR Text Key111986901
Report Number2249723-2018-01038
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/29/2018
Initial Date FDA Received06/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-