| Model Number |
5376 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that during device change-out procedure, atrial set screw was stuck in the header and atrial lead could not be removed.Further investigation revealed a piece of grommet which was slightly obscuring the set screw hole.Device was explanted and replaced.Patient¿s condition was stable.
|
| |
|