• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367297
Medical Device Problem Code Volume Accuracy Problem (1675)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/24/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
In this mdr, bd corporate headquarters in (b)(4) has been listed in as nipro is an oem manufacturing site.Bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode with the incident lot was not observed.A review of the manufacturing records was completed for the incident lot and no issues were identified.
 
Event or Problem Description
It was reported that bd vacutainer® safety-lok¿ blood collection sets had blood stop flowing through the tubing upon reaching the female luer hub.No serious injury or medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
MDR Report Key7618745
Report Number2243072-2018-01191
Device Sequence Number13918680
Product Code JKA
UDI-Device Identifier50382903672979
UDI-Public50382903672979
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K991088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2019
Device Catalogue Number367297
Device Lot Number6J1181
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2017
Initial Report FDA Received Date06/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-