• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 1818910 SIGMA HP TROCH ANT BLK SZ 4R; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 1818910 SIGMA HP TROCH ANT BLK SZ 4R; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES Back to Search Results
Catalog Number 202474400
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the surgeon was inserting a headed pin into the cut guide and it got stuck.No surgical delay.All pieces were retrieved.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned device confirmed the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP TROCH ANT BLK SZ 4R
Common Device Name
KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7618785
Report Number1818910-2018-62634
Device Sequence Number13858943
Product Code HWT
UDI-Device Identifier10603295083955
UDI-Public10603295083955
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number202474400
Device Lot NumberAH0609
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2018
Initial Date Received by Manufacturer 12/07/2017
Supplement Date Received by Manufacturer05/30/2018
Initial Report FDA Received Date06/20/2018
Supplement Report FDA Received Date06/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age38 YR
-
-