Model Number 18880
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Information (3190)
Date of Event 05/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the cuff started to leak after 12 days of use.Patient status was asked but unknown.
 
Additional Manufacturer Narrative
Evaluation summary: the failure cut in cuff was confirmed in the received sample.The device history record was not reviewed since the lot number was not provided.Manufacturing process was reviewed and currently, there is no potential risk and the below condition was found: an inflation/deflation test is performed for all taperguard products.The operator inspects all products for no leaks and segregates the defective parts.All taperguard products have a resting time of 2 hours.Once the part is inspected visually, then the instruction provides the guidelines to deflate the cuff.All manufacturing controls were found acceptable and effective to detect the reported failure mode.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALLINCKRODT
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ SA DE CV(USD)
ave henequen no 1181 desarrol
ciudad juarez,mx 32590
MX  32590
MDR Report Key7620123
Report Number2936999-2018-00402
Device Sequence Number1397282
Product Code BTR
UDI-Device Identifier30884521032549
UDI-Public30884521032549
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number18880
Device Catalogue Number18880
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2018
Initial Date Received by Manufacturer 06/05/2018
Supplement Date Received by Manufacturer09/14/2018
Initial Report FDA Received Date06/20/2018
Supplement Report FDA Received Date10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-