MPRI CAPSUREFIX NOVUS; ELECTRODE, PACEMAKER, PERMANENT Back to Search Results
Model Number 5076-85
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: ddbc3d1 icd, implanted (b)(6) 2013; 1688t-85 lead, implanted (b)(6) 2007.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead exhibited noise and oversensing was noted on some stored episodes.The patient underwent defibrillation thresholds testing, which was successful without complications.The lead remains in use and no changes were made.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing associated with the right ventricular lead.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS
Common Device Name
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7621118
Report Number2649622-2018-10274
Device Sequence Number1695328
Product Code DTB
UDI-Device Identifier00681490730099
UDI-Public00681490730099
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/06/2007
Device Model Number5076-85
Device Catalogue Number5076-85
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/31/2018
Supplement Date Received by Manufacturer05/31/2018
Initial Report FDA Received Date06/20/2018
Supplement Report FDA Received Date06/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
693352 LEAD, 6936110 LEAD
Patient Age69 YR
-
-