| Model Number |
15 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control then replaced the device's main keypad assembly and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.Physio-control further evaluated the removed main keypad assembly and determined that the cause of the reported issue was that the left side of the on button was worn.As a result, it required a very firm press in order for it to engage and power the device on.The right side of the button was ok.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device's on/off button was intermittently not responding when pressed.As a result, defibrillation could have been delayed, or prevented, if it were necessary.There was no patient use associated with the reported event.
|
| |
|