• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 05/29/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control then replaced the device's system pcb assembly and completed other, unrelated, repairs.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.
 
Event Description
The customer contacted physio-control to report that their device would not complete the boot-up process in order to power on.As a result, defibrillation would not have been possible if it were necessary.There was no patient use associated with the reported issue.
 
Manufacturer Narrative
Physio-control further evaluated the removed system pcb assembly and determined that the cause of the reported issue was that the single board computer (sbc) located on the assembly was inoperative.
 
Event Description
The customer contacted physio-control to report that their device would not complete the boot-up process in order to power on.As a result, defibrillation would not have been possible if it were necessary.There was no patient use associated with the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
meg marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7621354
MDR Text Key111850165
Report Number0003015876-2018-00971
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873988039
UDI-Public00883873988039
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15
Device Catalogue Number99577-000045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/06/2018
Initial Date FDA Received06/20/2018
Supplement Dates Manufacturer Received09/10/2018
Supplement Dates FDA Received09/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-