Model Number 15 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/29/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control then replaced the device's system pcb assembly and completed other, unrelated, repairs.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.
|
|
Event Description
|
The customer contacted physio-control to report that their device would not complete the boot-up process in order to power on.As a result, defibrillation would not have been possible if it were necessary.There was no patient use associated with the reported issue.
|
|
Manufacturer Narrative
|
Physio-control further evaluated the removed system pcb assembly and determined that the cause of the reported issue was that the single board computer (sbc) located on the assembly was inoperative.
|
|
Event Description
|
The customer contacted physio-control to report that their device would not complete the boot-up process in order to power on.As a result, defibrillation would not have been possible if it were necessary.There was no patient use associated with the reported issue.
|
|