Model Number SK13535M
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device was returned to the manufacturer for evaluation.The investigation is currently under way.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a treatment of a highly calcified lesion in sfa, the recanalization crossing support catheter allegedly broke in segments.It was further reported that a snare device was used to capture and retrieve all the segments.Reportedly, there was no further treatment.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a treatment of a highly calcified lesion in sfa, the recanalization crossing support catheter allegedly broke in segments.It was further reported that a snare device was used to capture and retrieve all the segments.Reportedly, there was no further treatment.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: one seeker catheter was returned for evaluation.The distal end of the returned device was noted to be stretched and jagged, indicating a detachment.The returned catheter was measured and it was determined 10.7cm of the catheter was not returned for evaluation.Therefore, the investigation is confirmed for the alleged detachment.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEEKER SUPPORT CATHETER
Common Device Name
SUPPORT CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7621390
Report Number2020394-2018-00927
Device Sequence Number2067199
Product Code DQY
UDI-Device Identifier00801741067891
UDI-Public(01)00801741067891
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K103367
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSK13535M
Device Catalogue NumberSK13535M
Device Lot NumberVTBS0074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2018
Initial Date Received by Manufacturer 05/24/2018
Supplement Date Received by Manufacturer08/15/2018
Initial Report FDA Received Date06/20/2018
Supplement Report FDA Received Date08/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
Patient Weight76
-
-