C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC SPIRIT SHEATH STYLE 3 WIDEBAND 36MM 100BX; MALE EXTERNAL CATHETER
|
Back to Search Results |
|
| Model Number |
39104 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
Pain (1994)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the adhesive on the male external catheters was too strong, and pulled at the skin of the patient's penis.The patient allegedly experienced soreness and pain during urination for several days after removal.No medical intervention was required.
|
| |
|
Event or Problem Description
|
|
It was reported that the adhesive on the male external catheters were too strong, and pulled at the skin of the patient's penis.The patient allegedly experienced soreness and pain during urination for several days after removal.No medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual evaluation noted 60 unopened samples in original packaging.No obvious defects.Adhesive peel test was performed.Samples were cut (0.6910, 0.6925, 0.7085, 0.7070, 0.6985, 0.6985, 0.7185, 0.7000, 0.7050, 0.7010, 0.6995, 0.6945, and 0.6945in) to the required width (0.70" +/- 0.05 ").Adhesive peel strength (0.8899, 0.9380, 0.9171, 0.9825, 0.8925, 1.0037, 1.0353, 1.0924, 1.0814, 1.2017, 0.9866, 0.9864, and 1.0808lbf) was found to be within specification (0.60-2.10lbf).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication: do not use on irritated or compromised skin.Precaution: do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply: wash penis with mild soap and warm water.Dry thoroughly.Trim pubic hair if necessary.Open package at perforation.To remove plastic insert, squeeze catheter at the top of the white cone and pull to release.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove.Gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual evaluation noted 60 unopened samples in original packaging.No obvious defects.Adhesive peel test was performed.Samples were cut (0.6910, 0.6925, 0.7085, 0.7070, 0.6985, 0.6985, 0.7185, 0.7000, 0.7050, 0.7010, 0.6995, 0.6945, and 0.6945in) to the required width (0.70" +/- 0.05 ").Adhesive peel strength (0.8899, 0.9380, 0.9171, 0.9825, 0.8925, 1.0037, 1.0353, 1.0924, 1.0814, 1.2017, 0.9866, 0.9864, and 1.0808lbf) was found to be within specification (0.60-2.10lbf).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) open package at perforation.4) to remove plastic insert, squeeze catheter at the top of the white cone and pull to release.5) unroll self-adhering catheter over penis.6) gently squeeze the catheter to properly seal adhesive to the skin.7) connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." correction: mfr site, concomitant medical products and device evaluated by mfr.
|
| |
|
Event or Problem Description
|
|
It was reported that the adhesive on the male external catheters were too strong, and pulled at the skin of the patient's penis.The patient allegedly experienced soreness and pain during urination for several days after removal.No medical intervention was required.
|
| |
|