• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC SPIRIT SHEATH STYLE 3 WIDEBAND 29MM 100BX; MALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC SPIRIT SHEATH STYLE 3 WIDEBAND 29MM 100BX; MALE EXTERNAL CATHETER Back to Search Results
Model Number 39102
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the patient's doctor allegedly believed that the male external catheters may have contributed to the patient developing a bladder infection, for which the patient was hospitalized.It was reported that the patient had been receiving antibiotics through a picc line every eight hours for two weeks to treat the bladder infection.It was reported that the patient discontinued use of the male external catheters after diagnosis of the bladder infection.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿description/indication: the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication: do not use on irritated or compromised skin.Precaution: do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions to apply: wash penis with mild soap and warm water.Dry thoroughly.Trim pubic hair if necessary.Open package at perforation.To remove plastic insert, squeeze catheter at the top of the white cone and pull to release.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions to remove: gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive".
 
Event or Problem Description
It was reported that the patient's doctor allegedly believed that the male external catheters may have contributed to the patient developing a bladder infection, for which the patient was hospitalized.It was reported that the patient received antibiotics through a picc line every eight hours for two weeks to treat the bladder infection.It was reported that the patient discontinued use of the male external catheters after diagnosis of the bladder infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC SPIRIT SHEATH STYLE 3 WIDEBAND 29MM 100BX
Common Device Name
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7621522
Report Number1018233-2018-02353
Device Sequence Number1397247
Product Code NNX
UDI-Device Identifier00801741071287
UDI-Public(01)00801741071287
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/28/2022
Device Model Number39102
Device Catalogue Number39102
Device Lot NumberJUBY0080
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/30/2018
Supplement Date Received by Manufacturer07/17/2018
Initial Report FDA Received Date06/20/2018
Supplement Report FDA Received Date07/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-