| Model Number |
FG-0201 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Uterine Perforation (2121)
|
| Date of Event |
05/31/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a symphion resecting device used in a hysteroscopy procedure performed on (b)(6) 2018.According to the complainant, during procedure and inside the patient, the resecting device was used to do an endometrial sampling and noticed that there was a 400cc fluid loss early in the case.The physician advanced a little further with good visualization but at some point the deficit jumped to 800cc.Upon procedure closure, the physician suspected perforation and a leak into the abdomen.A laparoscopy was done and a hole was found in the uterus with a little bit of fluid in the cavity.The physician stitched it up and suctioned any fluid out from the patient.Reportedly, the physician was unsure if the perforation was caused by the symphion resecting device or the dilation at the beginning of the case.The procedure was completed using this device.There were no patient complications as a result of the event.The patient condition at the conclusion of the procedure was reported to be fine.
|
| |
|