| Model Number |
72202674 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that surgeon was doing a meniscal repair using fastfix, knot pusher did not cut the suture.No backup device was readily available.No procedural delay or patient injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Codes added.
|
| |
|
Additional Manufacturer Narrative
|
|
One fast-fix 360 kpsc was returned for evaluation.Functional assessment of the device could not confirm the reported complaint of being dull.The device was tested using fast-fix ultrabraid suture and cut as intended.Per the device ifu "while holding the suture taut, gently slide the knot pusher/suture cutter along the suture to the knot to achieve the desired tension.Continuing to hold the suture taut, push the knot pusher/suture cutter tip against the knot to recess the knot to the meniscus, and cut the suture by sliding the trigger forward".
|
| |
|
Additional Manufacturer Narrative
|
|
Examination was not possible, as the device has not been returned.The investigation could not draw any conclusions about the reported event without the return of the device.
|
| |
|