• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC RETRACTA DETACHABLE EMBOLIZATION COIL; KRD DEVICE, EMBOLIZATION, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC RETRACTA DETACHABLE EMBOLIZATION COIL; KRD DEVICE, EMBOLIZATION, VASCULAR Back to Search Results
Model Number G34198
Health Effect - Clinical Code No Information (3190)
Date of Event 09/03/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
It was reported that during the varicocele embolization procedure, the physician felt resistance at the distal part of the catheter.The embolization coil needed extra force to advance; therefore, the user retracted the coil and noted a kink at the connection to the delivery wire.Another device was used to complete the procedure successfully.Furthermore, it was noted that the proximal end of the coil was stretched and stuck in the loading cartridge.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETRACTA DETACHABLE EMBOLIZATION COIL
Common Device Name
KRD DEVICE, EMBOLIZATION, VASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7621963
Report Number1820334-2018-01937
Device Sequence Number1898686
Product Code KRD
UDI-Device Identifier10827002341989
UDI-Public(01)10827002341989(17)180901(10)4475659
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2018
Device Model NumberG34198
Device Catalogue NumberMWCER-35-14-8
Device Lot Number4475659
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2018
Initial Report FDA Received Date06/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-