| Catalog Number |
MOB-10 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The implant fragments were examined visually using a light microscope.The fractured surface was homogeneous and shows no obvious root cause of a material failure.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018 a dentist from (b)(6) reported that a (b)(6) -year-old female patient had a 3m¿ espe¿ 2.4x10mm mdi o-ball prosthetic head - collared (mob-10) implant fractured.On (b)(6) 2018 it was reported to 3m that the implant was surgically removed without complication using a lindemann burr.The current patient status was reported to be fine.The implant was placed in the upper jaw in (b)(6) 2013.The reason for implant breakage could not be determined.
|
| |
|