It was happened that vit2 appeared on eva and the screen went red during vit surgery, and to clear the error they turned off, rebooted and did re-priming with infusion tubing connected with patient eye, because his nurse insisted that our young sales reps say it is possible by phone conversation.The priming completed successfully.After surgery, his patient lost the sight of his eye and requested him to submit his clinical record.The dr.Really would like to know possible impacts that might be caused by continuing to apply 100mmhg of irrigation pressure to the eye for at least 1 minute during priming.He would like to know the reason his patient lost the sight of his eye and if it was because of eva or not.The dr.Insists that the priming should not completed with infusion tubing connected with patient eye.This patient had an artery stenosis and further investigation determined that the artery stenosis was present prior to the start of the surgery.In collaboration with the physician we conclude that this is the main reason for the lost of sight by the patient.
|