MAKO SURGICAL CORP. ONLAY INSERT EXTRACTOR; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 160430
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker has initiated a voluntary, lot specific recall for the mako onlay insert extractor which has not been assigned an fda reference number.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.Not returned to the manufacturer.
 
Event or Problem Description
Rep reported that on sales branch advice, all sets were inspected and a second instrument was found rusted.No associated procedure.
 
Additional Manufacturer Narrative
An event regarding corrosion involving a mako extractor was reported.The device was discovered during inspection.The product was not returned for evaluation.There was no surgical procedure associated with the reported event.This event meets the definition of preventive maintenance; no further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Rep reported that on sales branch advice, all sets were inspected and a second instrument was found rusted.No associated procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONLAY INSERT EXTRACTOR
Common Device Name
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key7622337
Report Number3005985723-2018-00383
Device Sequence Number19187857
Product Code NPJ
UDI-Device Identifier00848486020644
UDI-Public00848486020644
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K090763
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 09/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number160430
Device Lot Number19110915
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/23/2018
Supplement Date Received by Manufacturer08/29/2018
Initial Report FDA Received Date06/20/2018
Supplement Report FDA Received Date09/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2051-2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-