• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX; VELA INFRARENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX; VELA INFRARENAL Back to Search Results
Model Number A28-28/C95V
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/10/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device was not returned as the device remains implanted therefore no evaluation was completed.At the completion of the clinical assessment, clinical was able to find substantial evidence to support the following of a type ib of the left common iliac artery.The most likely cause of the loss of seal could not be determined.There are no further reports of negative patient sequelae.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.
 
Event or Problem Description
At the completion of the clinical assessment on (b)(6) /2018 for a pervious report event (2031527-2017-00219), clinical was able to find substantial evidence to support the following additional findings of type ib endoleak of the left common iliac artery.This report is for the type ib endoleak only.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
VELA INFRARENAL
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key7623475
Report Number2031527-2018-00553
Device Sequence Number1898109
Product Code MIH
UDI-Device Identifier00818009013521
UDI-Public(01)00818009013521(17)151103
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/03/2015
Device Model NumberA28-28/C95V
Device Lot Number1252155-011
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/22/2018
Initial Report FDA Received Date06/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
-
-