BARD PERIPHERAL VASCULAR, INC. CONQUEST PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER
|
Back to Search Results |
|
Model Number CQ7568 |
Device Problems
Material Frayed (1262); Peeled/Delaminated (1454); Unraveled Material (1664)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/23/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 11/2020).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Event Description
|
It was reported that during an angioplasty procedure in the radial artery, upon removing the pta balloon from the patient, unraveled material was found.No further treatment was required.There was no reported patient injury.
|
|
Event Description
|
It was reported that during an angioplasty procedure in the radial artery, upon removing the pta balloon from the patient, unraveled material was found.No further treatment was required.There was no reported patient injury.
|
|
Manufacturer Narrative
|
The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.The device was returned for evaluation.A visual inspection found the outer layer to be peeled from the balloon throughout the length of the balloon.Additionally, unraveled fibers were noted on the distal cone of the balloon.Therefore, the investigation is confirmed for peeled outer layer and unraveled material.The definitive root cause for the identified peeling outer layer and unraveled fibers could not be determined based upon the available information.An investigation is currently open to address the peeling outer layer issue.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Expiry date: 11/2020).(b)(4).
|
|