BARD PERIPHERAL VASCULAR, INC. CONQUEST PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQ7568
Medical Device Problem Codes Material Frayed (1262); Peeled/Delaminated (1454); Unraveled Material (1664)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date:11/2020).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the radial artery, upon removing the pta balloon from the patient, unraveled material was found.No further treatment was required.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation.A visual inspection found the outer layer to be peeled from the balloon throughout the length of the balloon.Additionally, fiber disturbance was noted on the distal cone of the balloon.No unraveled fibers was noted to the balloon.Therefore, the investigation is confirmed for peeled outer layer and fiber disturbance, and is unconfirmed for unraveled material.The definitive root cause for the identified peeling outer layer and fiber disturbance could not be determined based upon the available information.An investigation is currently open to address the peeling outer layer issue.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date:11/2020).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the radial artery, upon removing the pta balloon from the patient, unraveled material was found.No further treatment was required.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7623519
Report Number2020394-2018-00940
Device Sequence Number11118472
Product Code DQY
UDI-Device Identifier00801741062919
UDI-Public(01)00801741062919
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQ7568
Device Catalogue NumberCQ7568
Device Lot NumberREBZ1010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2018
Initial Date Received by Manufacturer 05/24/2018
Supplement Date Received by Manufacturer09/06/2018
Initial Report FDA Received Date06/20/2018
Supplement Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-