Device is a combination product.(b)(4).Device evaluated by mfr: stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent found that the proximal end of the stent was damaged and bunched in a distal direction.The distal end of the stent was still in its crimped position.The undamaged crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and do not appear to have been subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the outer and the inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
Reportable based on additional information received on 01-jun-2018.It was reported that crossing difficulties were encountered.The 90% stenosed, 38mm x 3.5mm, de novo target lesion with a bend of between 45 and 90 degrees was located in the moderately tortuous and moderately calcified left anterior descending artery.Following pre-dilatation using a 2.0mm balloon catheter resulting to 70% residual stenosis, a 3.50 x 38 synergy¿ drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with another 3.50 x 38 synergy¿ drug-eluting stent.No patient complications were reported and the patient's status was stable.However, it was further reported that the stent was damaged during advancing.
|