| Model Number |
AAB00 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted; give date: n/a (not applicable).The intraocular lens was not implanted.If explanted; give date: n/a (not applicable).The intraocular lens was not implanted.Ring on the lens.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the intraocular lens (iol) had a ring on it.This was noticed during preparation, so there was no patient contact.No additional information was provided to johnson and johnson surgical vision, inc.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the returned sample was received in its original packaging.The product inspected using a 12x microscope magnification, showed deep scratches on the optic area and machine line observed outside of 3mm area, which is still acceptable.The complaint event cannot be confirmed since the machine line was still within the acceptable limit and how or when the scratches were introduced could not be determined.The complaint cannot be confirmed.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.There was no discrepancy found during the review.A search revealed that no other complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|