| Model Number |
75533 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250); Connection Problem (2900)
|
| Health Effect - Clinical Codes |
Blood Loss (2597); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation is in progress, but not yet concluded.(b)(4).A sample was received and decontaminated.Both visual and photographic review indicated that there was crazing on the female large bore luer, and a crack parallel to the parting line.The sample was also tested for leaks and was found to leak at a pressure of 20 mmhg.A representative sample was used to test the impact of solvent exposure to the connector, the test showed that no cracking on crazing occurred when exposed to solvent for 20+ hours.Another representative sample was used to test the impact of over tightening the luer when mated with the male counterpart.No cracking or crazing occurred as a result of this test.The sample will be sent to supplier for further evaluation.The device history record (dhr) was reviewed and there were no issues noted.The incoming raw material documentation, including the certificate of conformance and inspection results, were reviewed and there were no issues noted.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported that during cardiopulmonary bypass (cpb) procedure, the large luer on end of line (17b, l2) was a cracked and leaking.It resulted in approximately 5 cc blood loss.The device was changed out.The surgical procedure was completed successfully but there was a 5 minutes delay occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
The supplier inspected the part returned and confirmed the crack in the threaded part of the component as well as stress marks along the thread.They attempted to replicate the defect by overtightening the mated component and trying to crush the component with no success.They said that the root cause of the damage of the threaded section of the component is unknown at this time.The damage in the component is most likely caused by an incidental and isolated mold closure on the component during the molding process.There was no containment actions taken since they determined the issue to be an isolated incident (incident rate (b)(4)) and the lot has been depleted in total from their inventory.They did inspect current inventory and no issues were found.All available information has been placed on file in the quality management system for appropriate tracking, trending and follow-up.
|
| |
|