| Catalog Number |
KE152 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
05/24/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent kyphoplasty at l1 due to vertebral compression fracture.Intra-op, balloon ruptured and the marker got separated and was left in the vertebral body.The procedure was completed successfully with the original product.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to imdrf harmonization, some previously submitted device, method, result, and conclusion codes related to this event may have been updated.X-ray review: post-op x-rays for l1 kyphoplasty show a retained balloon marker.Balloon rupture can occur as a result of inflation over rated pressure or volume, or withdrawing a partially inflated balloon against the pedicle access needle.If information is provided in the future, a supplemental report will be issued.
|
| |
|