• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK KYPHON EXPRESS II INFLATABLE BONE TAMP; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK KYPHON EXPRESS II INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number KE152
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 05/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent kyphoplasty at l1 due to vertebral compression fracture.Intra-op, balloon ruptured and the marker got separated and was left in the vertebral body.The procedure was completed successfully with the original product.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Due to imdrf harmonization, some previously submitted device, method, result, and conclusion codes related to this event may have been updated.X-ray review: post-op x-rays for l1 kyphoplasty show a retained balloon marker.Balloon rupture can occur as a result of inflation over rated pressure or volume, or withdrawing a partially inflated balloon against the pedicle access needle.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON EXPRESS II INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
MDR Report Key7624839
Report Number1030489-2018-00879
Device Sequence Number10814076
Product Code HRX
UDI-Device Identifier00643169143777
UDI-Public00643169143777
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K123771
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberKE152
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/24/2018
Supplement Date Received by Manufacturer06/29/2018
08/08/2018
Initial Report FDA Received Date06/21/2018
Supplement Report FDA Received Date07/25/2018
10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age80 YR
Patient Weight89
-
-