• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 10 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 10 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-21-007
Medical Device Problem Codes Loss of or Failure to Bond (1068); Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Swelling (2091); Not Applicable (3189)
Date of Event 08/28/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Received medical records on 05/30/2018.Records reviewed for mdr reportability.Left attune knee revised to address pain, persistent swelling, and gross flexion instability.Surgical diagnosis was failed left total knee, secondary to instability, malrotation, and tibial base plate loosening.Revising surgeon observed synovitis, malrotation of the tibial tray, gross loosening/debonding of the tibial tray at the cement to implant interface, and loosening of the femur component at the cement to bone interface (femur and cement came loose together with simple taps of the mallet).The patella, and the tibia cement mantle, were noted to be well-fixed.Tibial insert was described as pitted, but not delaminated.Patella was not revised; otherwise, all other depuy products replaced with competitor products.Doi: (b)(6) 2015; dor: (b)(6) 2017; (left knee).
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #
=
> (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FB INSRT SZ 10 7MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key7625010
Report Number1818910-2018-62823
Device Sequence Number11333113
Product Code JWH
UDI-Device Identifier10603295046264
UDI-Public10603295046264
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2017
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Model Number1516-21-007
Device Catalogue Number151621007
Device Lot Number427515
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/30/2018
Supplement Date Received by Manufacturer12/11/2018
11/08/2019
05/07/2020
Initial Report FDA Received Date06/21/2018
Supplement Report FDA Received Date12/12/2018
11/13/2019
05/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-