MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 030453
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during low anterior resection surgery, while grasping rectum, the green button was pushed and enter in firing mode however the handle did not fire although the center control down was pushed multiple times.The surgical time was extended by less than 30 min.They opened new handle and it worked fine.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.There were no visual or functional abnormalities noted with the products received.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7625131
Report Number1219930-2018-03290
Device Sequence Number1799404
Product Code GDW
UDI-Device Identifier10884523002843
UDI-Public10884523002843
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number030453
Device Catalogue Number030453
Device Lot NumberN7E0998KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2018
Initial Date Received by Manufacturer 06/06/2018
Supplement Date Received by Manufacturer08/20/2018
Initial Report FDA Received Date06/21/2018
Supplement Report FDA Received Date09/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-