DAVOL INC., SUB. C.R. BARD, INC. X-STREAM LAPAROSCOPIC IRRIGATION CONTROLLER; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 5551000
Medical Device Problem Codes Thermal Decomposition of Device (1071); Failure to Power Up (1476); Use of Device Problem (1670)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation identified areas of physical damage on the outer casing.The void seal was broken, indicating the controller may have been opened.The unit was sold to the user in december of 2016.Based on the evaluation, we have concluded that the physical damage and the heat damage identified with the internal components was most likely due to the device being subjected to physical trauma.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported per the customer contact that during testing, the x-stream controller would not power on.There was no patient involvement.The subject product was returned and evaluation identified that some internal components had heat damage with evidence of internal combustion having occurred.There was no report of a short circuit, smoke, or flame having occurred during use of the device.It is not known when the internal damage occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-STREAM LAPAROSCOPIC IRRIGATION CONTROLLER
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC.
100 crossings blvd.
warwick RI 02886
MDR Report Key7625222
Report Number1213643-2018-02091
Device Sequence Number18876612
Product Code GCJ
UDI-Device Identifier00801741010118
UDI-Public(01)00801741010118
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K003790
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 06/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5551000
Device Lot NumberDAAVNX01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2018
Initial Report FDA Received Date06/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-