• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Codes Chest Pain (1776); Hyperglycemia (1905); Dizziness (2194)
Date of Event 05/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-18- user has high glucose value test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.Insurance nurse is calling on behalf of the customer.The customer is concerned with all of the results obtained.The expected fasting blood glucose test result range is unknown.The customer did report symptoms of feeling dizzy and had chest pain.Medical attention is reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the bedroom.During the call on (b)(6) 2018, a back to back blood test was not performed.The test strip lot manufacturer's expiration date is 04/23/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).Called customer back after his insurance agent called in to report that he sought medical intervention while using our meter.Customer states that on (b)(6) 2018, he was feeling dizzy and had chest pain.Customer states that he checked his blood sugar with our meter and the results was 308 mg/dl non-fasting, customer called the paramedic and they arrived within 10-minutes, they checked his blood sugar with their meter and the results was 402 mg/dl non-fasting.Customer was transported to the er and was diagnosed with chest pain.Customer was also treated for high blood sugar.Customer was released on (b)(6) 2018 with no new recommendation.Customer does not know his expected range.Customer states that he always tests non-fasting.Educated customer on proper way to test.Even though the comparison between the medic's meter and the metrix meter proves that the metrix is reading lower, customer states that he is not concern with low results, he is concerned with high results.Customer states that he feels well at the time of the call.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7625398
Report Number1000113657-2018-00671
Device Sequence Number1318085
Product Code NBW
UDI-Device Identifier00021292007911
UDI-Public(01)00021292007911
Combination Product (Y/N)N
Initial Reporter StateNJ
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 06/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMU2655
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2018
Initial Report FDA Received Date06/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY
Outcome Attributed to Adverse Event Hospitalization;
-
-