• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. AZUR 18 PERIPHERAL V-TRAK SYSTEM; EMBOLIZATION COIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. AZUR 18 PERIPHERAL V-TRAK SYSTEM; EMBOLIZATION COIL Back to Search Results
Model Number 45-480820
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history record indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The coil remains implanted in the patient and the remainder of the device was discarded at the user facility; therefore, a product evaluation could not be performed.The root cause is unknown.
 
Event or Problem Description
It was reported that during positioning of an embolization coil in a splenic artery, the coil detached in the microcatheter.The coil was pushed into the intended area with a guidewire.There was no reported intervention or patient injury.The patient was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR 18 PERIPHERAL V-TRAK SYSTEM
Common Device Name
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key7625488
Report Number2032493-2018-00117
Device Sequence Number1277446
Product Code HCG
UDI-Device Identifier00810170017988
UDI-Public(01)00810170017988(11)160701(17)210701(10)160701LC4
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K050954
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/01/2021
Device Model Number45-480820
Device Catalogue Number45-480820
Device Lot Number160701LC4
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/22/2018
Initial Report FDA Received Date06/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age46 YR
Patient Weight65
-
-