PENUMBRA, INC. PENUMBRA SYSTEM ACE 64 HI-FLOW KIT; NRY Back to Search Results
Catalog Number 5MAXACE064KIT
Medical Device Problem Codes Kinked (1339); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the ace64 was kinked approximately 59.0 cm from the hub.Bends were present along the length of the ace64.Conclusions: evaluation of the returned ace64 revealed a kinked device.This type of damage typically occurs due to improper handling during removal from packaging.If a device is forcefully retracted at extreme angles during removal from the packaging or otherwise mishandled during preparation, damage such as kinks may occur.No other devices associated with the complaint were returned for evaluation.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a thrombectomy procedure, the hospital technician noted that the penumbra system ace 64 reperfusion catheter (ace64) was kinked upon removal from the packaging of the penumbra system ace 64 hi-flow kit (kit).The damage to the ace64 was found prior to use and therefore, was not used in the procedure.The procedure was completed using a new ace64.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ACE 64 HI-FLOW KIT
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7625670
Report Number3005168196-2018-01220
Device Sequence Number1317462
Product Code NRY
UDI-Device Identifier00814548016467
UDI-Public00814548016467
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K160449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Radiologic Technologist
Device Expiration Date01/02/2021
Device Catalogue Number5MAXACE064KIT
Device Lot NumberF81263
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2018
Initial Report FDA Received Date06/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-