Patient information is unknown.Date of event is unknown.Device is an instrument and is not implanted/explanted.Address and other contact information is unknown.A device history record review was requested.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Reporter¿s phone number is unknown.A device history record (dhr) review was conducted for part number: 03.632.074, synthes lot number: 9824265: release to warehouse date: 19-oct-2015, manufacturing site: synthes (b)(4): no non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.A product investigation was conducted.The t25 stardrive screwdriver (part: 03.632.074, lot: 9824265, mfg: 19-oct-2015) was inspected at us customer quality (cq).Upon visual inspection, it was observed that the distal tip of the t25 stardrive screwdriver¿s shaft was worn and rounded/stripped.Also, the t25 stardrive screwdriver shows signs of minimal wear for its 2-year lifetime.No new malfunctions have been identified.The received condition agrees with the complaint description.Dimensional inspection: dimensional analysis was not performed due to post-manufacture damage of the small hexagonal pelvic screwdriver.Document/specification review: relevant drawing(s) was reviewed; no design or manufacturing defect or deficiency was observed during the investigation conclusion: the complaint condition is confirmed.While no definitive root cause could be determined, it is possible that the device encountered unintended forces (such as being dropped, struck off-axis or damaged during usage or handling or sterile processing).Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|