AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1550225-15
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.It should be noted that the xience sierra everolimus eluting coronary stent system instruction for use (ifu) states: when treating multiple lesions within the same vessel, stent the distal lesion prior to stenting the proximal lesion.Stenting in this order obviates the need to cross the proximal stent during placement of the distal stent, and reduces the chance of damaging or dislodging the proximal stent.It appears the ifu deviation may have contributed to the reported event.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely the device interacted with the previously implanted stent while crossing distally causing the reported physical resistance and subsequent stent dislodgement with the patient effect of foreign body in patient.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the chronic total occlusion in the right coronary artery.Two xience sierra stents were successfully implanted in the mid to proximal right coronary artery (rca).The 2.25 x 15 mm xience sierra was advance into the patient anatomy, to treat a diseased portion of the vessel that was distal to the two implanted stents.Resistance was noted advancing through the previously implanted stents; however, the device was able to cross to the target lesion.Once at the target lesion, it was noted that the stent was not on the delivery system.The stent was not seen in the patient anatomy; however, it was felt that the stent was on the delivery system when it was inserted into the patient anatomy.The stent could not be located outside the patient anatomy.A new xience sierra stent was then used to treat the remaining lesion.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7626036
Report Number2024168-2018-04774
Device Sequence Number1201351
Product Code NIQ
UDI-Device Identifier08717648227011
UDI-Public08717648227011
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/11/2019
Device Catalogue Number1550225-15
Device Lot Number8032641
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2018
Initial Report FDA Received Date06/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
XIENCE SIERRA STENTS: 2.25X38, 2.25X28MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
-
-