• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 05/09/2018
Event Type  malfunction  
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and replaced the hkr processor board to resolve the error.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The replaced board was requested for return to livanova (b)(4) for further investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that an s5 double head pump displayed an error message during priming.There was no patient involvement.
 
Manufacturer Narrative
The replaced hkr processor board was returned to livanova (b)(4) for further investigation.During evaluation of the returned device, the reported failure was reproduced.The root cause was determined to be errors in the internal memory bank in the micro-controller.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7626047
MDR Text Key111999537
Report Number9611109-2018-01059
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817900405
UDI-Public010403381790040511150819
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/10/2018
Initial Date FDA Received06/21/2018
Supplement Dates Manufacturer Received06/25/2018
Supplement Dates FDA Received07/25/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-