Catalog Number RBY2C0825
Medical Device Problem Code Kinked (1339)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, the distal end of a ruby coil pusher assembly became kinked as the hospital staff was advancing it out of the introducer sheath and into a lantern delivery microcatheter (lantern).The ruby coil was, therefore, removed, and the procedure was completed using a new ruby coil.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The embolization coil was intact with the pusher assembly and the coil was undamaged.The pusher assembly was kinked approximately 20.0 cm, 39.0 cm, 57.0 cm, 78.0 cm, 105.0 cm and 107.0 cm from the proximal end.The coil was able to be advanced out of its introducer sheath and through a demonstration microcatheter without any issue.Conclusions: evaluation of the returned device revealed there were kinks along the length of the pusher assembly.If the device is forcefully advanced against resistance, the pusher assembly may become kinked.Further damage to the pusher assembly may have occurred during packaging for return to penumbra.During functional testing, the returned ruby coil was able to be advanced through a demonstration microcatheter without an issue.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7626320
Report Number3005168196-2018-01222
Device Sequence Number13903066
Product Code HCG
UDI-Device Identifier00814548012971
UDI-Public00814548012971
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/23/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2026
Device Catalogue NumberRBY2C0825
Device Lot NumberF82260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 05/23/2018
Supplement Date Received by Manufacturer10/19/2018
Initial Report FDA Received Date06/21/2018
Supplement Report FDA Received Date10/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Date Report Sent to FDA01/01/2005
-
-