Model Number 15
Device Problem Failure To Run On AC/DC (1001)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/31/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device.The reported issue was both verified and duplicated.Physio-control replaced the battery pins and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.
 
Event Description
The customer contacted physio-control to report that their device display went blank when attempting to print a code summary.This behavior is indicative of a device lockup condition and, as a result, defibrillation therapy would not be possible.Physio-control has made multiple attempts to contact the customer in order to obtain additional information about both the patient and the event; however, no response has been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7627363
MDR Text Key111984936
Report Number0003015876-2018-00992
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15
Device Catalogue Number99577-000046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Initial Date Manufacturer Received 06/04/2018
Initial Date FDA Received06/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-