• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERILMED, INC. MULTIPLE CLIP APPLIER LIGACLIP MCA; CLIP, IMPLANTABLE, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERILMED, INC. MULTIPLE CLIP APPLIER LIGACLIP MCA; CLIP, IMPLANTABLE, REPROCESSED Back to Search Results
Catalog Number ETHMSM20
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 01/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The sterilmed failure analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.(b)(4).
 
Event or Problem Description
It was reported that a patient underwent a procedure with a clip appliers ligaclip multiple clip appliers lg clips and the clips did not fire correctly which caused the vessel to be severed.During the procedure, during use of the device the damage of the vessel was identified immediately.The patient¿s vessel was repaired and extended hospitalization was not required.The patient has since fully recovered.The physician¿s opinion on the cause of the adverse event is that it was due to product malfunction.
 
Additional Manufacturer Narrative
The account returned 1 ethmsm20 stating that, "clips did not fire correctly" and that during the procedure the vessel was severed.The device was examined and subjected to full visual and functional testing.Outside of the presence of contaminants consistent with patient contact and use in the field, the device passed its visual inspection.The jaws were in alignment and the clip retainer was in position.The functionality of the device was tested via firing the device.The device was shipped to the field with 17 clips for use as noted by the number written in ink on the handle.The device was returned with 8 clips remaining.The device fired all 8 remaining clips, producing clips off appropriate pinch and alignment without interruption until all clips were spent.No defect was displayed by the device, nor were any of the clips found to displaying any features that would result in an injury.Based on this evidence the account's complaint could not be duplicated nor confirmed.A device history record review was performed, and no discrepancies were noted.Manufacturer's ref.No: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIPLE CLIP APPLIER LIGACLIP MCA
Common Device Name
CLIP, IMPLANTABLE, REPROCESSED
Manufacturer (Section D)
STERILMED, INC.
5010 cheshire parkway
ste 2
plymouth MN 55446
MDR Report Key7627397
Report Number2134070-2018-00104
Device Sequence Number1818364
Product Code NMJ
UDI-Device Identifier10888551019095
UDI-Public10888551019095
Combination Product (Y/N)N
PMA/510(K) Number
K033579
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberETHMSM20
Device Lot Number2037507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2018
Initial Date Received by Manufacturer 05/29/2018
Supplement Date Received by Manufacturer05/29/2018
Initial Report FDA Received Date06/21/2018
Supplement Report FDA Received Date07/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-