• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE MICROFX OCD UNIVERSAL DRILL; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE MICROFX OCD UNIVERSAL DRILL; BIT, DRILL Back to Search Results
Catalog Number 234-200-200
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available, it will be provided in future reports.
 
Event or Problem Description
It was reported that the product broke during use.All broken pieces were retrieved and the procedure was completed successfully.
 
Additional Manufacturer Narrative
Alleged failure: microfracture was being carried out in a knee arthroscopy procedure using the microfx instrumentation.The microfx ocd drill bit broke at the distal end into to 2 pieces, leaving other small bits of metal debris within the joint.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be: excessive force applied on drill, running drill in reverse, drill is started/stopped while in bone, or hard bone.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.The device manufacturer date is not known.
 
Event or Problem Description
It was reported that the product broke during use.All broken pieces were retrieved and the procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROFX OCD UNIVERSAL DRILL
Common Device Name
BIT, DRILL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key7627612
Report Number0002936485-2018-00498
Device Sequence Number1819018
Product Code HTW
UDI-Device Identifier37613327004091
UDI-Public37613327004091
Combination Product (Y/N)N
Initial Reporter StateVI
Initial Reporter CountryAU
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number234-200-200
Device Lot NumberLOT 64776
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2018
Initial Date Received by Manufacturer 05/24/2018
Supplement Date Received by Manufacturer05/24/2018
Initial Report FDA Received Date06/21/2018
Supplement Report FDA Received Date07/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-