• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED CORPORATION 4.5MM X 19CM HPS SPHERICAL BUR; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED CORPORATION 4.5MM X 19CM HPS SPHERICAL BUR; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number HPS-H001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The broken hps-h001 is not expected to be returned for evaluation and review.This complaint of broken device was verified via photographs that were received.The photographs were reviewed by manufacturing engineering and it was determined that the breakage was caused by excessive lateral force.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.There have been a total of 10 complaints for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of (b)(4) complaints, regarding 27 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use for this accessory's handpiece, the user is advised the following: precautions: do no apply excessive loading on the shaver blade or bur.Excessive force or using shaver blades or burs as a lever can cause damage to the device including permanent deformation, shedding of metal (wear), motor seizure, and overheating.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
The sales representative reported on behalf of the customer that the hps-h001, 4.5 mm x 19 cm hps spherical bur, device broken mid-shaft during surgery.It was reported that the device fragment/component was retrieved from the patient.The patient did not sustain any injury from this event.Further information has been requested; however, that information has not been made available.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM X 19CM HPS SPHERICAL BUR
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
CONMED CORPORATION
11311 concept blvd.
largo FL 33773
MDR Report Key7628386
Report Number1017294-2018-00090
Device Sequence Number16738210
Product Code GFA
UDI-Device Identifier20845854020682
UDI-Public(01)20845854020682(17)230123(10)898656
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date01/23/2023
Device Catalogue NumberHPS-H001
Device Lot Number898656
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2018
Initial Report FDA Received Date06/22/2018
Patient Sequence Number1
-
-