• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-712 PARADIGM INSULIN PUMP; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-712 PARADIGM INSULIN PUMP; PUMP, INFUSION, INSULIN Back to Search Results
Model Number MMT-712EWS
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/30/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer reported via phone call that the insulin pump had a keypad anomaly and the act button was unresponsive.Customer¿s blood glucose was unknown at the time of incident.No troubleshooting was performed.Insulin pump is being replaced and is expected to return for analysis.
 
Manufacturer Narrative
Device received with intermittent button response due to flattened dome switch on act button.Connector was inspected and locked on lcd board.No button error alarm during testing.Device passed displacement test and self test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MMT-712 PARADIGM INSULIN PUMP
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7628570
MDR Text Key112219353
Report Number3004209178-2018-84270
Device Sequence Number1
Product Code LZG
UDI-Device Identifier00613994469694
UDI-Public(01)00613994469694
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-712EWS
Device Catalogue NumberMMT-712EWS
Device Lot NumberA9712EWSJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received12/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/21/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-