COOK ENDOSCOPY 6 SHOOTER SAEED MULTI-BAND LIGATOR; MND, LIGATOR, ESOPHAGEAL Back to Search Results
Catalog Number MBL-6-I-5B-S
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report of premature band deployment prior to use.The product received was returned in an open pouch.The return included the trigger cord attached to the barrel with four (4) bands, one (1) amber band and three (3) black bands.The two (2) deployed black bands were not included in the return.A crack along the length of the barrel was also observed.The length of the trigger cord measured within tolerance.It is unknown how or at what point the bands deployed in the packaging or the barrel was cracked.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual product preparation and handling conditions could not be duplicated in the laboratory setting.This limits our ability to conclusively determine a cause.Premature band deployment and damage to the barrel can occur if the device experiences excessive pressure during general handling or shipment.Prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
In preparation for a procedure, the user selected a cook 6 shooter saeed multi-band ligator.The customer opened the cook band [ligator] and the ring [band] was already displaced from its place (premature band deployment).This occurred prior to patient contact; there was no impact to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6 SHOOTER SAEED MULTI-BAND LIGATOR
Common Device Name
MND, LIGATOR, ESOPHAGEAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key7628708
Report Number1037905-2018-00269
Device Sequence Number19427155
Product Code MND
UDI-Device Identifier00827002268616
UDI-Public(01)00827002268616(17)180530(10)W3870550
Combination Product (Y/N)N
Initial Reporter CountryIN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2018
Device Catalogue NumberMBL-6-I-5B-S
Device Lot NumberW3870550
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2018
Initial Report FDA Received Date06/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-