| Model Number |
4593 |
| Medical Device Problem Codes |
Capturing Problem (2891); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Event or Problem Description
|
|
Boston scientific received information that this left ventricular (lv) lead had slightly elevated threshold that was noted during device changeout.Boston scientific technical services (ts) discussed options with pacing system analyzer (psa).Additional information received indicating that high threshold was possibly due to micro dislodgement but was not confirmed.There was no action taken.This lv lead remains in service.No adverse patient effects were reported.
|
| |
|