| Model Number |
MMT-522NAS |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump had prime rewind and insulin squirted out if the tubing during manual prime.The customer¿s blood glucose was unknown.Customer states they did not receive 2nd series of beeps nor did numbers appear on the screen.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Motor error alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to confirm prime or fill anomaly due to motor error alarm noted.
|
| |
|