U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number RLT231416
Medical Device Problem Code Collapse (1099)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Additional device: (b)(4).It is not known which device appeared to have become compressed, therefore both devices were investigated.A review of the manufacturing records for the devices verified the lots met all pre-release specifications.According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), in the pre-treatment planning it states: determine accurate size of anatomy and proper size of trunk-ipsilateral and contralateral endoprostheses (tables 21 and 22) and aortic and iliac extender endoprostheses (tables 23 and 24).Per the trunk-ipsilateral leg endoprosthesis sizing guide, the intended iliac vessel diameter for the rlt23416 device is 12-13.5 mm.Per the contralateral leg endoprosthesis sizing guide the intended iliac vessel diameter for the plc271000 21.5-25 mm.
 
Event or Problem Description
The following information was reported to gore: on (b)(6) 2018 a patient was undergoing treatment of a common iliac artery aneurysm with gore® excluder® aaa endoprostheses and gore® excluder® iliac branch endoprostheses.The patient did not have an abdominal aortic aneurysm, and the gore® excluder® iliac branch component was seated entirely in the common iliac artery.A 27 mm contralateral leg component was used as a bridge between the contralateral gate and the iliac branch component.The distal aorta appears to be measuring approximately 20 mm.On (b)(6) 2018 it was recognized that one of the device limbs of the gore® excluder® aaa endoprostheses appeared to have been compressed.On (b)(6) 2018 a reintervention was performed whereby a 14 mm begraft stent was placed within the non-compressed limb and a 12 mm begraft stent was placed on the compressed side.The patient did well following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key7629771
Report Number3007284313-2018-00186
Device Sequence Number1432860
Product Code MIH
UDI-Device Identifier00733132622382
UDI-Public00733132622382
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/14/2019
Device Catalogue NumberRLT231416
Device Lot Number15692950
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/22/2018
Supplement Report FDA Received Date06/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient Age66 YR
-
-