It was reported that a (b)(6) female patient underwent a ventricular tachycardia (vt) ablation procedure for premature ventricular contractions (pvcs) with a thermocool® sf nav bi-directional catheter and suffered a cardiac tamponade requiring pericardiocentesis.Pre-procedure, prior to any catheters being inserted, a small pericardial effusion was detected via intracardiac echocardiography (ice).Post-procedure, during the waiting period, and prior to catheter removal, the patient became hypotensive and the effusion was confirmed to have increased via ice.Pericardiocentesis yielded an unspecified volume of fluid.Blood pressure normalized.Patient was reported to be in stable condition and was transported to the intensive care unit (icu) for observation.There is no information regarding extended hospitalization.Factors cited that may have contributed to the adverse event include the small pericardial effusion identified prior to catheter introduction.Physician¿s opinion regarding the cause of the adverse event is that although a small pericardial effusion was identified prior to ablation, the physician believed that the adverse event was caused by hyperdynamic cardiac wall motion secondary to ablation and the use of isuprel.There were no issues, errors, or malfunctions reported on any bwi products or equipment during the procedure.No transseptal puncture was performed, as a retrograde aortic approach was used to access the left ventricle.There is no sheath information.Generator was set on power control mode at 20-30 watts with temperature cut-off at 40 degrees celsius and impedance spike cut-off of 250 ohms.Temperature and impedance remained within normal limits throughout the procedure.Power was not titrated.There is no information regarding overall ablation time at the site of injury or last ablation cycle time at the site of injury, as the site of injury was not identified.There is no information regarding irrigated catheter flow setting.Patient received anticoagulant during the procedure with activated clotting time (act) goal of 300-350 seconds.
|